University of Sussex
Browse
sensors-21-00999 (2).pdf (3.92 MB)

Characterisation of textile embedded electrodes for use in a neonatal smart mattress electrocardiography system

Download (3.92 MB)
Heart rate monitoring is the predominant quantitative health indicator of a newborn in the delivery room. A rapid and accurate heart rate measurement is vital during the first minutes after birth. Clinical recommendations suggest that electrocardiogram (ECG) monitoring should be widely adopted in the neonatal intensive care unit to reduce infant mortality and improve long term health outcomes in births that require intervention. Novel non-contact electrocardiogram sensors can reduce the time from birth to heart rate reading as well as providing unobtrusive and continuous monitoring during intervention. In this work we report the design and development of a solution to provide high resolution, real time electrocardiogram data to the clinicians within the delivery room using non-contact electric potential sensors embedded in a neonatal intensive care unit mattress. A real-time high-resolution electrocardiogram acquisition solution based on a low power embedded system was developed and textile embedded electrodes were fabricated and characterised. Proof of concept tests were carried out on simulated and human cardiac signals, producing electrocardiograms suitable for the calculation of heart rate having an accuracy within ±1 beat per minute using a test ECG signal, ECG recordings from a human volunteer with a correlation coefficient of ~ 87% proved accurate beat to beat morphology reproduction of the waveform without morphological alterations and a time from application to heart rate display below 6 s. This provides evidence that flexible non-contact textile-based electrodes can be embedded in wearable devices for assisting births through heart rate monitoring and serves as a proof of concept for a complete neonate electrocardiogram monitoring system.

History

Publication status

  • Published

File Version

  • Published version

Journal

Sensors

ISSN

1424-8220

Publisher

MDPI

Issue

3

Volume

21

Page range

1-20

Article number

a999

Department affiliated with

  • Clinical and Experimental Medicine Publications

Full text available

  • Yes

Peer reviewed?

  • Yes

Legacy Posted Date

2021-02-03

First Open Access (FOA) Date

2021-02-03

First Compliant Deposit (FCD) Date

2021-02-03

Usage metrics

    University of Sussex (Publications)

    Categories

    No categories selected

    Licence

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM
    DC