A multicenter measurement of magnetization transfer ratio in normal white matter

Berry, Isabelle, Barker, Gareth J., Barkhof, Frederik, Campi, Adriana, Dousset, Vincent, Franconi, Jean-Michel, Gass, Achim, Schreiber, Wolfgang, Miller, David H. and Tofts, Paul S. (1999) A multicenter measurement of magnetization transfer ratio in normal white matter. Journal of Magnetic Resonance Imaging, 9 (3). pp. 441-446. ISSN 1053-1807

Full text not available from this repository.

Abstract

To assess the importance of intercenter variations when measuring magnetization transfer ratio (MTR) in the brain, six European centers measured MTR in normal white matter. MTR ranged from 9 to 51 percent units (25 sequences). The effective flip angle of the saturating pulse divided by the pulse repetition time (ENRsat degrees/msec) was a good predictor of MTR (MTR = 3.25 ENRsat).

Item Type: Article
Keywords: Brain, Anatomy & histology, Metabolism, Magnetic Resonance Imaging, Nerve Fibers, Myelinated, Reproducibility of Results, Sensitivity and Specificity
Schools and Departments: Brighton and Sussex Medical School > Brighton and Sussex Medical School
Depositing User: Paul Stephen Tofts
Date Deposited: 28 Feb 2007
Last Modified: 03 Sep 2019 12:52
URI: http://sro.sussex.ac.uk/id/eprint/826
Google Scholar:63 Citations