An Ultra Low Noise Electric Potential Probe for Human Body Scanning
journal contribution
posted on 2023-06-08, 00:18 authored by Robert Prance, A Debray, T D Clark, H Prance, M Nock, C J Harland, A J ClippingdaleIn this paper we describe a new very-low-noise, high-input-impedance probe developed to make non-contact measurements of electrical potentials generated by currents flowing in the human body. With a noise level of 2 µV Hz-1/2 at 1 Hz, down to 0.1 µV Hz-1/2 at 1 kHz, and an operational bandwidth from 0.01 Hz to 100 KHz, this probe would seem well suited to the detection of a wide range of electrical activity in the body.
History
Publication status
- Published
Journal
Measurement Science and TechnologyISSN
09570233Publisher
Measurement Science and TechnologyExternal DOI
Issue
3Volume
11Page range
291-297ISBN
0957-0233Department affiliated with