A multicenter measurement of magnetization transfer ratio in normal white matter

Berry, I., Barker, G. J., Barkhof, F., Campi, A., Dousset, V., Franconi, J. M., Gass, A., Schreiber, W., Miller, D. H. and Tofts, P. S. (1999) A multicenter measurement of magnetization transfer ratio in normal white matter. J Magn Reson Imaging, 9 (3). pp. 441-6. ISSN 1053-1807 (Print)

Full text not available from this repository.

Abstract

To assess the importance of intercenter variations when measuring magnetization transfer ratio (MTR) in the brain, six European centers measured MTR in normal white matter. MTR ranged from 9 to 51 percent units (25 sequences). The effective flip angle of the saturating pulse divided by the pulse repetition time (ENRsat degrees/msec) was a good predictor of MTR (MTR = 3.25 ENRsat).

Item Type: Article
Uncontrolled Keywords: Brain/ anatomy & histology/metabolism Humans Linear Models Magnetic Resonance Imaging/ methods Nerve Fibers, Myelinated/ metabolism Reference Values Reproducibility of Results Sensitivity and Specificity
Divisions: (3502) School of Business, Management and Economics > (817) Brighton and Sussex Medical School
Depositing User: Prof Paul Stephen Tofts
Date Deposited: 28 Feb 2007
Last Modified: 08 Dec 2010 14:48
URI: http://sro.sussex.ac.uk/id/eprint/826

Actions (login required)