Berry, I., Barker, G. J., Barkhof, F., Campi, A., Dousset, V., Franconi, J. M., Gass, A., Schreiber, W., Miller, D. H. and Tofts, P. S. (1999) A multicenter measurement of magnetization transfer ratio in normal white matter. J Magn Reson Imaging, 9 (3). pp. 441-6. ISSN 1053-1807 (Print)
Full text not available from this repository.Official URL: http://dx.doi.org/10.1002/(SICI)1522-2586(199903)9...
Abstract
To assess the importance of intercenter variations when measuring magnetization transfer ratio (MTR) in the brain, six European centers measured MTR in normal white matter. MTR ranged from 9 to 51 percent units (25 sequences). The effective flip angle of the saturating pulse divided by the pulse repetition time (ENRsat degrees/msec) was a good predictor of MTR (MTR = 3.25 ENRsat).
| Item Type: | Article |
|---|---|
| Uncontrolled Keywords: | Brain/ anatomy & histology/metabolism Humans Linear Models Magnetic Resonance Imaging/ methods Nerve Fibers, Myelinated/ metabolism Reference Values Reproducibility of Results Sensitivity and Specificity |
| Divisions: | (3502) School of Business, Management and Economics > (817) Brighton and Sussex Medical School |
| Depositing User: | Prof Paul Stephen Tofts |
| Date Deposited: | 28 Feb 2007 |
| Last Modified: | 08 Dec 2010 14:48 |
| URI: | http://sro.sussex.ac.uk/id/eprint/826 |